• BOD Water Quality Online Analyzer
BOD Water Quality Online Analyzer
  • Applicable Medium
    Water
  • Not applicable medium
    Gas
  • Product Introduction
    Biochemical oxygen demand (BOD) is mainly used to monitor the pollution status of organic matter in water bodies. Generally, organic matter can be decomposed by microorganisms, but microorganisms require oxygen to decompose organic compounds in water. If the dissolved oxygen in the water is not sufficient to meet the needs of microorganisms, the water body will be in a polluted state.

Working principle

Heating a mixture of water sample, potassium dichromate digestion solution, silver sulfate solution (silver sulfate as a catalyst can effectively oxidize linear fatty compounds), and concentrated sulfuric acid to 170 ℃ will cause a color change after the oxidation of organic matter in the solution by dichromate ions. The analyzer detects this color change and converts it into BOD value output. The amount of dichromate ions consumed is equivalent to the amount of oxidizable organic matter.


The reducing inorganic substances in water samples, such as nitrite, sulfide, and ferrous ions, will react with potassium dichromate, affecting the measurement results. The amount of potassium dichromate consumed by them will be recorded in the measurement results, making the measurement results higher. The interference of chloride ions in water samples can be eliminated by adding mercuric sulfate, as chloride ions can form very stable mercuric chloride with mercury ions.


Product Features

1. Select valve components: Choose reagent sampling timing, flexible and diverse channels, versatile functions, easy maintenance, high lifespan, and other advantages.

2. Micro metering component: achieves precise measurement of reagents through a visual photoelectric system, overcoming quantitative errors caused by wear of peristaltic pump tubes; At the same time, precise quantification of trace reagents has been achieved, with each dose being only 1 milliliter, greatly reducing the amount of reagents used.

3. Sample injection component: peristaltic pump negative pressure suction, there is always an air buffer zone between the reagent and the pump tube to avoid corrosion of the pump tube; At the same time, it makes the mixing of reagents more concise and flexible.

4. Sealed digestion component: High temperature and high pressure digestion system accelerates the reaction process and overcomes the corrosion of equipment caused by the volatilization of corrosive gases in open systems.

5. Reagent tube: Imported modified PTFE transparent hose with a diameter greater than 1.5mm is used to reduce the probability of particle blockage in water samples.

Technical Parameter

model

YK-DH319BD

measuring method

Dichromate method

measuring range

10~2000   mg/L BOD

Accuracy

Not exceeding ± 10%

stability

Not exceeding ± 10%

Sample making cycle

The minimum measurement cycle is 40 minutes

Measurement cycle

Time interval (adjustable from 40 to 9999 minutes) and hourly measurement mode

Maintenance cycle

Usually once a month, about 30 minutes each time

Calibration cycle

Adjustable at any interval and time from 0 to 99 days

output

RS232 or RS485; 4-20mA output (expandable); Relay output

Switching quantity

1-6 relay outputs can be used, with a capacity of 220VAC/2A

Environmental requirements

Suggested temperature+5-40 ℃; Humidity ≤ 85% (non condensing)

size

Height 1430 x Width 500 x Depth 403 (mm)

Power Supply

AC220±10%V,50±10%Hz,5A

other

Touch screen display and command input; 