This instrument uses potassium permanganate (acidic) titration method to measure water samples. Dissolve and oxidize certain organic and inorganic reducing substances in the water sample using an acidic potassium permanganate solution at 96 ℃, and measure the remaining amount of potassium permanganate at 525nm. The reduction of potassium permanganate is directly proportional to the consumption of organic matter.
Modular design allows each component to operate independently, improving the efficiency of the instrument;
• Various types of signal outputs: printers RS485/RS232、4-20mA、GPRS、 Relays, etc;
Various sample making methods: manual control, timed (interval and hourly) control, 4-20mA current input control, switch control, and other forms;
Automatic compensation for measuring light source attenuation, good stability, and small drift;
Wide range of measurement, fast sampling speed, and high accuracy;
Low dosage of medication, maintenance free for one month;
Optoelectronic non-contact measurement, high measurement accuracy, and high operational reliability;
Adopting high-resolution industrial grade color touch screen, user-friendly interface, configuration design, and simple operation.
application area
Online monitoring of water quality at the inlet and outlet of municipal sewage treatment plants
Online monitoring of groundwater and surface water quality
Online monitoring of drinking water quality
Online monitoring of river, lake, reservoir, and seawater quality
Testing principle: Potassium permanganate (acidic) titration method
Reference method: GB/T 11892-89
Dissolving temperature: 96 ℃
Dissolving time: 1000s
Measurement range: 0-60mg/L (expandable)
Inspection basis: HJ/T 100-2003
Accuracy: ± 5%
Repetitive error: ± 5%
Zero drift: ± 5%
Range drift: ± 5%
Colorimetric wavelength: 525nm
Minimum cycle: 40min
Analog output: 1 4-20mA output (optional 2 channels), maximum load resistance of 750 Ω
Digital output: RS232/RS485
Switch output: 1 relay output (optional 2)
Other outputs: Printer output or sending SMS to the target phone number via GPRS to create samples
Data storage: It can store measurement data for three years and can be stored cyclically
Data export: Measurement values can be imported into a USB drive through a USB port and saved
Power supply: AC220 ± 10% V, 50 ± 10% Hz, 5A
Dimensions: Height 1430 x Width 500 x Depth 403 (mm)